Product Description

SAR Measurement System(ESSAY-4)

MODEL : ESSAY-4
Robot Body
제품 사양
Product name Robot Body Quantity (pcs) 1
Manufacturer Adept(USA) Model name S850
Item See specifications
Robot type Vertical articulated robot
Degrees of freedom 6
Payload 5 kg
Iterative precision ± 0.03 mm
Arm length 854 mm
Operating area & maximum speed Axis 1 ± 170° (250°/sec)
Axis 2 -190°, +45° (250°/sec)
Axis 3 -29, +256° (250°/sec)
Axis 4 ± 190° (375°/sec)
Axis 5 ± 120° (375°/sec)
Axis 6 ± 360° (600°/sec)
Body weight 29 kg
Installation
environment
Ambient environment 5 ~ +40°C
Ambient humidity 5 % ~ 90 % R.H.
Power 24 VDC, 120 W
Robot stand Height 100 cm
Feature The robot table was made 100 cm high for greater use convenience (so that the user can measure the SAR without having to bend forward).
Adept(USA)
제품 사양
Product nam Robot Controller Quantity (pcs) 1
Manufacturer Adept(USA) Model name Smart Controller CX
Item

Emergency stop function: There is a separate manual controller for controlling the robot manually; the body of the controller and the separate manual controller include an emergency stop button.

Manual controller (Teaching Pendant): The manual controller can operate each of the six axes of the robot individually and supports the operation according to the spatial coordinates of the X, Y, and Z axes.

Can detect the opening/closing signals of the phantom cover to prevent collision between the phantom cover and the probes

Features

When the probe is mistakenly moved toward the covered phantom, it may be damaged. To prevent this, the system was designed to be capable of sensing the phantom cover open/close signals, thereby preventing damage to the probe.

When the probe is subjected to an impact, the system will stop its operation.

When there is a request from the user for robot control, the same can be immediately activated.

Electric-Field Probe
제품 사양
Product name Electric-Field Probe Quantity (pcs) 1
Manufacturer EMF Safety Co., Ltd. (South Korea) Model name ET25, ET10
규격

Form: 3-axis isotropic electric-field probe

Electric-field probe (1 pc): Frequency range 150 MHz–3 GHz

Electric-field probe (1 pc): Frequency range 150 MHz–6 GHz

Directivity: ± 0.2 dB in HSL (rotation on the probe axis)
                   ± 0.3 dB in tissue material(rotation normal to probe axis)

Linearity : ± 0.2 dB

Operating range: 0.01–100 W/kg

Measures each component of the electric field

Made of a material that can withstand even organic solvents

Calibration data indicating probe sensitivity, isotropy, etc. are stored and supplied in a portable storage device.

Calibration reports for this probe are provided.

A case is provided for storage and movement of the probe.

Features

The probe is the core part of the SAR measurement system and is manufactured and calibrated by our company.

Data-Processing Unit
제품 사양
Product name Data-Processing Unit Quantity (set) 1
Manufacturer EMF Safety Co., Ltd. (South Korea) Model name ES-DRS
Specifications

Functions: Amplifies the electrical signals received from the SAR measurement probe, transmits data using an optical cable, carries out A/D conversion, and senses any impact on the probe

Item Characteristics
Proximity sensor Probe protection sensor and emergency stop
Input Offset Automatic Zero
Battery 9 V battery (usable for more than 50 hours)
Features

The probe is the core part of the SAR measurement system and is manufactured and calibrated by our company.

Dipole Antenna for Effectiveness Verification
제품 사양
Product name Dipole Antenna for Effectiveness Verification Quantity (pcs) 12
Manufacturer EMF Safety Co., Ltd. (South Korea) Model name ES-DA-XXXXX
Specifications

The standard dipole antenna copies the output of the same frequency band as that of the measuring device instead of the measurement target device before starting SAR measurement to verify the effectiveness of the electric-field probe and data collection system for improving the accuracy of the measurement data.

Frequency [MHz] System test error SAR target value
300, 450, 835, 900, 1,800, 1,900, 1,950, 2100, 2,450, 3500, 5,200, 5800MHz ±10% of the target value

Type: Dipole antenna Connector: SAM or N type Input impedance: 50 Ω

Reflection loss s: > 20 dB under the phantom solution

Power performance:

> 100 W (f < 1 GHz), > 40 W (f > 1 GHz)

Includes a spacer for effectiveness evaluation

Includes an antenna-mounting table for antenna installation

Includes a case for storing and moving the dipole antenna

Features

As the antenna and the mounting table are separate, antenna installation is easy and simple, and antenna storage and management are easy.

Phantom Solution
제품 사양
Product name Phantom Solution Quantity (set) 12
Manufacturer EMF Safety Co., Ltd. (South Korea) Model name ES-HL-XXXXX
규격

15 phantom solutions for each measurement frequency band

Volume: 300/450/835/900/1,800/1,900/1,950/2,100/2,450/3,500/5,200/5,800 MHz

Tolerance range of electrical characteristics: Within ± 5%

Submits data from the measurement of permittivity and conductivity

Features

When a customer requests a phantom solution in a specific frequency band, it can be manufactured and supplied after consultation with the customer.